Recall Z-1498-2022— Siemens Healthcare Diagnostics, Inc.

Class II · Ongoing

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2022-06-28. It names one FDA 510(k) clearance: K991576. Product: Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative determination of iron in human serum and (lithium heparin) plasma samples.. Reason: Falsely Elevated Atellica CH Microalbumin_2 (¿ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately after an iron test..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Ongoing
Initiated
2022-06-28
Firm location
Tarrytown, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative determination of iron in human serum and (lithium heparin) plasma samples.

Reason for recall

Falsely Elevated Atellica CH Microalbumin_2 (¿ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately after an iron test.

Data sourced from openFDA. This site is unofficial and independent of the FDA.