Recall Z-1497-2026— Ion Beam Applications S.A.
Class II · Ongoing
Class II FDA medical device recall by Ion Beam Applications S.A., initiated 2026-02-03. It names one FDA 510(k) clearance: K163500. Product: IBA Proton Therapy System - PROTEUS 235. Reason: It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used..
- Recalling firm
- Ion Beam Applications S.A.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-02-03
- Firm location
- Ottignies-Louvain-La-Neuve, N/A, Belgium
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
IBA Proton Therapy System - PROTEUS 235
Reason for recall
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
Data sourced from openFDA. This site is unofficial and independent of the FDA.