Recall Z-1497-2023— Beaver Visitec International, Inc.

Class II · Ongoing

Class II FDA medical device recall by Beaver Visitec International, Inc., initiated 2023-03-17. It names one FDA 510(k) clearance: K202038. Product: CryoTreQ- Aa disposable handheld instrument intended for ophthalmic surgery. It creates a tip at cryogenic temperatures by evaporation of N2O that can be utilized to perform cryotherapy based on the destruction of tissue by extreme cold to locally perform cryocoagulation on tissue. Ref: CT00.D01. Reason: Device suddenly and forcefully disassembled after cryotherapy, potential for long-term or permanent impairment of a patient or user if a CryoTreq device suddenly or forcefully disassembles during a cryosurgery procedure..

Recalling firm
Beaver Visitec International, Inc.
Classification
Class II
Status
Ongoing
Initiated
2023-03-17
Firm location
Waltham, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

CryoTreQ- Aa disposable handheld instrument intended for ophthalmic surgery. It creates a tip at cryogenic temperatures by evaporation of N2O that can be utilized to perform cryotherapy based on the destruction of tissue by extreme cold to locally perform cryocoagulation on tissue. Ref: CT00.D01

Reason for recall

Device suddenly and forcefully disassembled after cryotherapy, potential for long-term or permanent impairment of a patient or user if a CryoTreq device suddenly or forcefully disassembles during a cryosurgery procedure.