Recall Z-1496-2017— Merge Healthcare, Inc.
Class II · Terminated
Class II FDA medical device recall by Merge Healthcare, Inc., initiated 2015-09-22, terminated 2017-04-18. It names one FDA 510(k) clearance: K051649. Product: Merge Cardio software with Issuer of Patient ID (IPID). The firm name on the label is Merge Healthcare, Hartland, WI.. Reason: For sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have the first name, last name, and medical record number in common..
- Recalling firm
- Merge Healthcare, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-09-22
- Terminated
- 2017-04-18
- Firm location
- Hartland, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Merge Cardio software with Issuer of Patient ID (IPID). The firm name on the label is Merge Healthcare, Hartland, WI.
Reason for recall
For sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have the first name, last name, and medical record number in common.
Data sourced from openFDA. This site is unofficial and independent of the FDA.