Recall Z-1492-2013— Stryker Instruments Div. of Stryker Corporation

Class II · Terminated

Class II FDA medical device recall by Stryker Instruments Div. of Stryker Corporation, initiated 2013-05-23, terminated 2014-04-14. It names one FDA 510(k) clearance: K070078. Product: Stryker Instruments0400-750-000 T4 Toga, Pullover, Regular Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.. Reason: Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr.

Recalling firm
Stryker Instruments Div. of Stryker Corporation
Classification
Class II
Status
Terminated
Initiated
2013-05-23
Terminated
2014-04-14
Firm location
Portage, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Stryker Instruments0400-750-000 T4 Toga, Pullover, Regular Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.

Reason for recall

Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

Data sourced from openFDA. This site is unofficial and independent of the FDA.