Recall Z-1490-2021— Boston Scientific Corporation

Class II · Ongoing

Class II FDA medical device recall by Boston Scientific Corporation, initiated 2021-04-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Colpassist Vaginal Positioning Device - Product Usage: intended for use in general gynecological surgery to assist in the positioning and manipulation of the vagina during the procedure. Material No. M0068318210. Reason: Potential for pinholes (sterile barrier breach) on the edge of the pouch to potentially render the device non-sterile, and may result in a post-operative infection.

Recalling firm
Boston Scientific Corporation
Classification
Class II
Status
Ongoing
Initiated
2021-04-15
Firm location
Marlborough, MA, United States

Product description

Colpassist Vaginal Positioning Device - Product Usage: intended for use in general gynecological surgery to assist in the positioning and manipulation of the vagina during the procedure. Material No. M0068318210

Reason for recall

Potential for pinholes (sterile barrier breach) on the edge of the pouch to potentially render the device non-sterile, and may result in a post-operative infection