Recall Z-1489-2022— Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Class I · Ongoing

Class I FDA medical device recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), initiated 2022-06-22. FDA's recall record names no specific 510(k) clearance for this event. Product: Crome Implantable Cardioverter Defibrillators: a. Crome VR ICD, Model Numbers: DVPC3D1, DVPC3D4; b. Crome DR ICD, Model Numbers: DDPC3D1, DDPC3D4. Reason: There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds)..

Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Classification
Class I
Status
Ongoing
Initiated
2022-06-22
Firm location
Mounds View, MN, United States

Product description

Crome Implantable Cardioverter Defibrillators: a. Crome VR ICD, Model Numbers: DVPC3D1, DVPC3D4; b. Crome DR ICD, Model Numbers: DDPC3D1, DDPC3D4

Reason for recall

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).