Recall Z-1489-2021— Ortho-Clinical Diagnostics, INc.
Class II · Terminated
Class II FDA medical device recall by Ortho-Clinical Diagnostics, INc., initiated 2021-03-26, terminated 2023-05-24. It names one FDA 510(k) clearance: K182063. Product: VITROS¿ Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part/ tool used in the reflectometer correction factor adjustment procedure.. Reason: Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride. If the customer doesnt have any White Reference Slides in reserve, they cant run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered. This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride..
- Recalling firm
- Ortho-Clinical Diagnostics, INc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-03-26
- Terminated
- 2023-05-24
- Firm location
- Rochester, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
VITROS¿ Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part/ tool used in the reflectometer correction factor adjustment procedure.
Reason for recall
Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride. If the customer doesnt have any White Reference Slides in reserve, they cant run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered. This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride.
Data sourced from openFDA. This site is unofficial and independent of the FDA.