Recall Z-1486-2022— Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class I · Ongoing
Class I FDA medical device recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), initiated 2022-06-22. FDA's recall record names no specific 510(k) clearance for this event. Product: Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt XT HF CRT-D, Model Numbers: DTPA2D4, DTPA2D1; b. Cobalt XT HF Quad CRT-D, Model Numbers: DTPA2QQ, DTPA2Q1. Reason: There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds)..
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2022-06-22
- Firm location
- Mounds View, MN, United States
Product description
Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt XT HF CRT-D, Model Numbers: DTPA2D4, DTPA2D1; b. Cobalt XT HF Quad CRT-D, Model Numbers: DTPA2QQ, DTPA2Q1
Reason for recall
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
Data sourced from openFDA. This site is unofficial and independent of the FDA.