Recall Z-1485-2013— Zimmer Surgical Inc

Class II · Terminated

Class II FDA medical device recall by Zimmer Surgical Inc, initiated 2013-05-07, terminated 2014-12-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Zimmer Air Dermatome II, REF 00-8851-001-01 handpiece, w/o hose, Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities.. Reason: This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site..

Recalling firm
Zimmer Surgical Inc
Classification
Class II
Status
Terminated
Initiated
2013-05-07
Terminated
2014-12-15
Firm location
Dover, OH, United States

Product description

Zimmer Air Dermatome II, REF 00-8851-001-01 handpiece, w/o hose, Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities.

Reason for recall

This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.