Recall Z-1484-2026— Siemens Healthcare Diagnostics, Inc.
Class II · Ongoing
Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2026-01-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Atellica CH Urine Albumin (UAlb). Material Number: 11537225. Reason: Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL..
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-01-20
- Firm location
- Tarrytown, NY, United States
Product description
Atellica CH Urine Albumin (UAlb). Material Number: 11537225
Reason for recall
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.