Recall Z-1483-2017— Acumed LLC

Class II · Terminated

Class II FDA medical device recall by Acumed LLC, initiated 2017-02-17, terminated 2017-05-19. FDA's recall record names no specific 510(k) clearance for this event. Product: .045 X 6 ST Guide Wire, Part number 45-0029-S, Sterile 10 pack of WS-1106ST and has a 6 inch length and 0.045 inch diameter. .054 X 7 ST Guide Wire, Part number 45-0031-S, Sterile 10 Pack of WS-1407ST and has a 7 inch length and 0.054 inch diameter. These guide wires are an orthopedic manual surgical instrument.. Reason: Guide wires sized as .045 X 6 were mis-packaged as .054 X 7 guide wires and those sized as .054 X 7 were mis- packaged as .045 X 6 guide wires. There is a possibility that using the incorrect guide wire size can cause health hazards, such as mal or nonunion, need for revision surgery, and histological response (if wire breaks)..

Recalling firm
Acumed LLC
Classification
Class II
Status
Terminated
Initiated
2017-02-17
Terminated
2017-05-19
Firm location
Hillsboro, OR, United States

Product description

.045 X 6 ST Guide Wire, Part number 45-0029-S, Sterile 10 pack of WS-1106ST and has a 6 inch length and 0.045 inch diameter. .054 X 7 ST Guide Wire, Part number 45-0031-S, Sterile 10 Pack of WS-1407ST and has a 7 inch length and 0.054 inch diameter. These guide wires are an orthopedic manual surgical instrument.

Reason for recall

Guide wires sized as .045 X 6 were mis-packaged as .054 X 7 guide wires and those sized as .054 X 7 were mis- packaged as .045 X 6 guide wires. There is a possibility that using the incorrect guide wire size can cause health hazards, such as mal or nonunion, need for revision surgery, and histological response (if wire breaks).

Data sourced from openFDA. This site is unofficial and independent of the FDA.