Recall Z-1481-2014— RAYSEARCH LABORATORIES AB
Class II · Terminated
Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2014-02-14, terminated 2017-02-15. It names 3 FDA 510(k) clearances: K130617, K120387, K100552. Product: RaySearch Ray Station Radiation Therapy Treatment Planning System, Stand-alone software. Versions 2.0, 2.5, 3.0, 3.5 and 4.0.. Reason: RaySearch Laboratories has recalled "RaySearch Ray Station 2.0, 2.5, 3.0, 3.5 and 4.0" software due to dose miscalculations for bolus/external/fixation support structures on CT-image..
- Recalling firm
- RAYSEARCH LABORATORIES AB
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-02-14
- Terminated
- 2017-02-15
- Firm location
- Stockholm, N/A, Sweden
510(k) clearances named in this recall
3 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
RaySearch Ray Station Radiation Therapy Treatment Planning System, Stand-alone software. Versions 2.0, 2.5, 3.0, 3.5 and 4.0.
Reason for recall
RaySearch Laboratories has recalled "RaySearch Ray Station 2.0, 2.5, 3.0, 3.5 and 4.0" software due to dose miscalculations for bolus/external/fixation support structures on CT-image.