Recall Z-1480-2026— Staar Surgical AG

Class II · Ongoing

Class II FDA medical device recall by Staar Surgical AG, initiated 2026-01-21. FDA's recall record names no specific 510(k) clearance for this event. Product: EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5. Reason: Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal..

Recalling firm
Staar Surgical AG
Classification
Class II
Status
Ongoing
Initiated
2026-01-21
Firm location
Nidau, N/A, Switzerland

Product description

EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Reason for recall

Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.