Recall Z-1480-2024— Abbott GmbH

Class II · Ongoing

Class II FDA medical device recall by Abbott GmbH, initiated 2024-03-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Alinity i Anti-HCV Reagent Kit, List Number 08P0521. Reason: Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05)..

Recalling firm
Abbott GmbH
Classification
Class II
Status
Ongoing
Initiated
2024-03-04
Firm location
Wiesbaden, Germany

Product description

Alinity i Anti-HCV Reagent Kit, List Number 08P0521

Reason for recall

Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).

Data sourced from openFDA. This site is unofficial and independent of the FDA.