Recall Z-1480-2024— Abbott GmbH
Class II · Ongoing
Class II FDA medical device recall by Abbott GmbH, initiated 2024-03-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Alinity i Anti-HCV Reagent Kit, List Number 08P0521. Reason: Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05)..
- Recalling firm
- Abbott GmbH
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-03-04
- Firm location
- Wiesbaden, Germany
Product description
Alinity i Anti-HCV Reagent Kit, List Number 08P0521
Reason for recall
Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).