Recall Z-1480-2021— Smiths Medical ASD Inc.
Class II · Terminated
Class II FDA medical device recall by Smiths Medical ASD Inc., initiated 2021-03-31, terminated 2023-03-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Tubing with male luer lock adapter (Component) a. 76 cm, REF 601280 b. 122 cm, REF 607281. component of extravascular blood-pressure Transducer.. Reason: Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices..
- Recalling firm
- Smiths Medical ASD Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-03-31
- Terminated
- 2023-03-20
- Firm location
- Minneapolis, MN, United States
Product description
Tubing with male luer lock adapter (Component) a. 76 cm, REF 601280 b. 122 cm, REF 607281. component of extravascular blood-pressure Transducer.
Reason for recall
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.