Recall Z-1479-2014— Hospira Inc.

Class II · Terminated

Class II FDA medical device recall by Hospira Inc., initiated 2014-02-25, terminated 2016-11-28. It names one FDA 510(k) clearance: K982159. Product: LifeShield LATEX-FREE HemoSet 100 mL Burette I.V. PlumSet, Nonvented, 105 inch with Prepierced Reseals, OPTION-LOK and Inline Dual Channel Cassette. For administration of blood and blood bags for use with Plum Series Infusers.. Reason: Hospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a burette and blood filter, intended for infusion of blood/blood products. In a gravity delivery, the correct lower lid dispenses 15 drops per mL and the incorrect lower lid found dispenses 10 drops per mL..

Recalling firm
Hospira Inc.
Classification
Class II
Status
Terminated
Initiated
2014-02-25
Terminated
2016-11-28
Firm location
Lake Forest, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

LifeShield LATEX-FREE HemoSet 100 mL Burette I.V. PlumSet, Nonvented, 105 inch with Prepierced Reseals, OPTION-LOK and Inline Dual Channel Cassette. For administration of blood and blood bags for use with Plum Series Infusers.

Reason for recall

Hospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a burette and blood filter, intended for infusion of blood/blood products. In a gravity delivery, the correct lower lid dispenses 15 drops per mL and the incorrect lower lid found dispenses 10 drops per mL.

Data sourced from openFDA. This site is unofficial and independent of the FDA.