Recall Z-1477-2023— Philips North America Llc
Class II · Ongoing
Class II FDA medical device recall by Philips North America Llc, initiated 2022-10-21. It names 2 FDA 510(k) clearances: K212837, K173433. Product: Philips ProxiDiagnost N90 1.0 (Model 706100) and ProxiDiagnost N90 1.1 (Model 706110)- A multi-functional general radiography and fluoroscopy (R/F) system. Reason: Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the user may not be able to initiate table tilting movements or braking.
- Recalling firm
- Philips North America Llc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-10-21
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips ProxiDiagnost N90 1.0 (Model 706100) and ProxiDiagnost N90 1.1 (Model 706110)- A multi-functional general radiography and fluoroscopy (R/F) system
Reason for recall
Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the user may not be able to initiate table tilting movements or braking