Recall Z-1476-2013— Hill-Rom, Inc.

Class II · Terminated

Class II FDA medical device recall by Hill-Rom, Inc., initiated 2013-04-30, terminated 2014-11-28. FDA's recall record names no specific 510(k) clearance for this event. Product: SlingBar Standard. Designed to meet the needs for lifting humans.. Reason: Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design..

Recalling firm
Hill-Rom, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-04-30
Terminated
2014-11-28
Firm location
Batesville, IN, United States

Product description

SlingBar Standard. Designed to meet the needs for lifting humans.

Reason for recall

Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design.

Data sourced from openFDA. This site is unofficial and independent of the FDA.