Recall Z-1472-2016— Customed, Inc

Class II · Terminated

Class II FDA medical device recall by Customed, Inc, initiated 2016-01-07, terminated 2017-05-03. FDA's recall record names no specific 510(k) clearance for this event. Product: Catalog Number: 900484 LAPAROSCOPIC GYNECOLOGY SURGICAL PACK Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.. Reason: Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile..

Recalling firm
Customed, Inc
Classification
Class II
Status
Terminated
Initiated
2016-01-07
Terminated
2017-05-03
Firm location
Fajardo, PR, United States

Product description

Catalog Number: 900484 LAPAROSCOPIC GYNECOLOGY SURGICAL PACK Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.

Reason for recall

Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.

Data sourced from openFDA. This site is unofficial and independent of the FDA.