Recall Z-1469-2021— Heartware, Inc.
Class II · Ongoing
Class II FDA medical device recall by Heartware, Inc., initiated 2021-02-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU. Heartware HVAD controller kits Models 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1420, and 1420JP. Heartware HVAD controller AC adapter Models 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, and 1430US. Heartware HVAD controller DC Adapter Model 1440. Heartware HVAD data cable - monitor Model 1575. Heartware HVAD battery pack Models 1650, 1650CA-CLIN, and 1650DE. Heartware HVAD patient pack Model 1475. Heartware HVAD waist pack Models 2050, 2050IL, and 2050US. Heartware HVAD shoulder pack Models 2060, 2060IL, and 2060US.. Reason: IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very short period may mean there is an intermittent problem and the cause of the alarm can be evaluated by analysis of the controller log files. The Patient Manual (PM) is being updated to instruct patients to contact their clinicians if they cannot identify an alarm before it is cleared as it might mean there is an intermittent problem. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87604.
- Recalling firm
- Heartware, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-02-26
- Firm location
- Miami Lakes, FL, United States
Product description
Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU. Heartware HVAD controller kits Models 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1420, and 1420JP. Heartware HVAD controller AC adapter Models 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, and 1430US. Heartware HVAD controller DC Adapter Model 1440. Heartware HVAD data cable - monitor Model 1575. Heartware HVAD battery pack Models 1650, 1650CA-CLIN, and 1650DE. Heartware HVAD patient pack Model 1475. Heartware HVAD waist pack Models 2050, 2050IL, and 2050US. Heartware HVAD shoulder pack Models 2060, 2060IL, and 2060US.
Reason for recall
IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very short period may mean there is an intermittent problem and the cause of the alarm can be evaluated by analysis of the controller log files. The Patient Manual (PM) is being updated to instruct patients to contact their clinicians if they cannot identify an alarm before it is cleared as it might mean there is an intermittent problem. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87604
Data sourced from openFDA. This site is unofficial and independent of the FDA.