Recall Z-1469-2020— The Binding Site Group, Ltd.

Class II · Terminated

Class II FDA medical device recall by The Binding Site Group, Ltd., initiated 2018-08-12, terminated 2022-01-20. FDA's recall record names no specific 510(k) clearance for this event. Product: VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016 - Product Usage: This assay is designed for the in vitro measurement of specific IgG antibodies against Haemophilus influenzae type b (Hib) capsular polysaccharide, present in human serum. Sufficient materials are supplied to allow a maximum of 41 samples to be tested in duplicate, with a calibration curve and two controls.. Reason: A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency..

Recalling firm
The Binding Site Group, Ltd.
Classification
Class II
Status
Terminated
Initiated
2018-08-12
Terminated
2022-01-20
Firm location
Birmingham, United Kingdom

Product description

VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016 - Product Usage: This assay is designed for the in vitro measurement of specific IgG antibodies against Haemophilus influenzae type b (Hib) capsular polysaccharide, present in human serum. Sufficient materials are supplied to allow a maximum of 41 samples to be tested in duplicate, with a calibration curve and two controls.

Reason for recall

A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.

Data sourced from openFDA. This site is unofficial and independent of the FDA.