Recall Z-1468-2026— Medline Industries, LP

Class II · Ongoing

Class II FDA medical device recall by Medline Industries, LP, initiated 2026-01-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167. Reason: Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body..

Recalling firm
Medline Industries, LP
Classification
Class II
Status
Ongoing
Initiated
2026-01-08
Firm location
Northfield, IL, United States

Product description

Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167

Reason for recall

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Data sourced from openFDA. This site is unofficial and independent of the FDA.