Recall Z-1463-2020— RAYSEARCH LABORATORIES AB

Class II · Terminated

Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2020-02-20, terminated 2024-11-05. It names one FDA 510(k) clearance: K190387. Product: RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system. Reason: There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect..

Recalling firm
RAYSEARCH LABORATORIES AB
Classification
Class II
Status
Terminated
Initiated
2020-02-20
Terminated
2024-11-05
Firm location
Stockholm, Sweden

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system

Reason for recall

There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.

Data sourced from openFDA. This site is unofficial and independent of the FDA.