Recall Z-1462-2024— Agfa N.V.
Class II · Ongoing
Class II FDA medical device recall by Agfa N.V., initiated 2024-02-23. It names one FDA 510(k) clearance: K103597. Product: Agfa DX-D 100 system-Digital Radiography mobile X-ray System. A mobile X-Ray modality, consisting of X-Ray console, generator, tube, digital detector, and workstation. This modality is used to generate projection X-Ray images of human patients. Models: 5411/ 050, A5411/ 0300 and A5411/ 0400. Reason: Steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is cut off does not work properly, may result in patient injury.
- Recalling firm
- Agfa N.V.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-02-23
- Firm location
- Mortsel, Belgium
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Agfa DX-D 100 system-Digital Radiography mobile X-ray System. A mobile X-Ray modality, consisting of X-Ray console, generator, tube, digital detector, and workstation. This modality is used to generate projection X-Ray images of human patients. Models: 5411/ 050, A5411/ 0300 and A5411/ 0400
Reason for recall
Steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is cut off does not work properly, may result in patient injury