Recall Z-1461-2014— Philips Medical Systems (Cleveland) Inc
Class II · Terminated
Class II FDA medical device recall by Philips Medical Systems (Cleveland) Inc, initiated 2014-03-27, terminated 2016-02-02. It names 4 FDA 510(k) clearances: K060937, K033357, K033326 and 1 more. Product: Brilliance CT 16, Brilliance CT 64, Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology, Brilliance iCT, Ingenuity CT, Ingenuity Core, Ingenuity Core 128 and Ingenuity Flex scanners are Computed Tomography X-ray systems.. Reason: Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe..
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-03-27
- Terminated
- 2016-02-02
- Firm location
- Cleveland, OH, United States
510(k) clearances named in this recall
4 on FDA's record- K060937 — BRILLIANCE VOLUME
- K033357 — ACQSIM-MULTISLICE-CT, MODEL 453567042271
- K033326 — PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM
- K012009 — MX8000 IDT CT SYSTEM
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Brilliance CT 16, Brilliance CT 64, Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology, Brilliance iCT, Ingenuity CT, Ingenuity Core, Ingenuity Core 128 and Ingenuity Flex scanners are Computed Tomography X-ray systems.
Reason for recall
Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.