Recall Z-1461-2013— Symbios Medical Products, LLC

Class I · Terminated

Class I FDA medical device recall by Symbios Medical Products, LLC, initiated 2013-05-10, terminated 2014-03-07. FDA's recall record names no specific 510(k) clearance for this event. Product: Part 510613 Disposable Infusion Pump, ...Elastomeric Pump 300mL (150 mL side)( 6mL/hr ),... found in the following kits GoBlock Kit 510612, Disposable Infusion Pump for drug delivery. Symbios GOBlock System is a disposable, self contained infusion system.. Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the.

Recalling firm
Symbios Medical Products, LLC
Classification
Class I
Status
Terminated
Initiated
2013-05-10
Terminated
2014-03-07
Firm location
Indianapolis, IN, United States

Product description

Part 510613 Disposable Infusion Pump, ...Elastomeric Pump 300mL (150 mL side)( 6mL/hr ),... found in the following kits GoBlock Kit 510612, Disposable Infusion Pump for drug delivery. Symbios GOBlock System is a disposable, self contained infusion system.

Reason for recall

Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

Data sourced from openFDA. This site is unofficial and independent of the FDA.