Recall Z-1455-2014— Gambro Renal Products, Incorporated

Class II · Terminated

Class II FDA medical device recall by Gambro Renal Products, Incorporated, initiated 2013-12-09, terminated 2015-12-22. FDA's recall record names no specific 510(k) clearance for this event. Product: Gambro Cartridge Blood Set, blood transport system for hemodialysis, Model Number(s): 101025 (003410510) - Cartridge set, STND PRM LN and 103401 (003414500) - Cartridge set, PRM_LN_INJ_PT. The Gambro Cartridge Sets are single use sterile tubing sets intended to provide extracorporeal blood transport circuit for hemodialysis treatments for Gambro Phoenix and COBE Centrisystem 3 (and 3+) Dialysis Delivery Systems.. Reason: Occluded heparin tubing events preventing anticoagulation dosing on the Gambro Cartridge Blood Sets..

Recalling firm
Gambro Renal Products, Incorporated
Classification
Class II
Status
Terminated
Initiated
2013-12-09
Terminated
2015-12-22
Firm location
Lakewood, CO, United States

Product description

Gambro Cartridge Blood Set, blood transport system for hemodialysis, Model Number(s): 101025 (003410510) - Cartridge set, STND PRM LN and 103401 (003414500) - Cartridge set, PRM_LN_INJ_PT. The Gambro Cartridge Sets are single use sterile tubing sets intended to provide extracorporeal blood transport circuit for hemodialysis treatments for Gambro Phoenix and COBE Centrisystem 3 (and 3+) Dialysis Delivery Systems.

Reason for recall

Occluded heparin tubing events preventing anticoagulation dosing on the Gambro Cartridge Blood Sets.

Data sourced from openFDA. This site is unofficial and independent of the FDA.