Recall Z-1453-2018— Siemens Medical Solutions USA, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2018-02-15, terminated 2020-04-08. It names one FDA 510(k) clearance: K050297. Product: SOMATOM Sensation 10 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.. Reason: to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure.

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Terminated
Initiated
2018-02-15
Terminated
2020-04-08
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

SOMATOM Sensation 10 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Data sourced from openFDA. This site is unofficial and independent of the FDA.