Recall Z-1453-2013— Symbios Medical Products, LLC
Class I · Terminated
Class I FDA medical device recall by Symbios Medical Products, LLC, initiated 2013-05-10, terminated 2014-03-07. FDA's recall record names no specific 510(k) clearance for this event. Product: Part 510076 Disposable Dual Infusion Pump, ...Elastomeric Dual Pump 300 mL(150 ml/side) 2mL/hr/side,... Sterile, found in the following kits: GoPump kit 510080-BP, 5 fenestrated catheter & BIOPATCH; GoPump kit 510112-BP, 2.5" fenestrated catheter & BIOPATCH Go Pump Kit 510141-BP,10" fenestrated catheter& BIOPATCH Disposable Single Infusion Pump for drug delivery Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication.. Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the.
- Recalling firm
- Symbios Medical Products, LLC
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2013-05-10
- Terminated
- 2014-03-07
- Firm location
- Indianapolis, IN, United States
Product description
Part 510076 Disposable Dual Infusion Pump, ...Elastomeric Dual Pump 300 mL(150 ml/side) 2mL/hr/side,... Sterile, found in the following kits: GoPump kit 510080-BP, 5 fenestrated catheter & BIOPATCH; GoPump kit 510112-BP, 2.5" fenestrated catheter & BIOPATCH Go Pump Kit 510141-BP,10" fenestrated catheter& BIOPATCH Disposable Single Infusion Pump for drug delivery Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication.
Reason for recall
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the
Data sourced from openFDA. This site is unofficial and independent of the FDA.