Recall Z-1452-2014— Draeger Medical, Inc.

Class II · Terminated

Class II FDA medical device recall by Draeger Medical, Inc., initiated 2014-03-25, terminated 2014-10-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Reusable latex breathing bags with multiple part numbers and sold as part of test lung part number and reusable anaesthesia sets. Product Usage: The breathing bag is a component of the anesthesia set (breathing circuit) that is intended to is the conduit for administering medical gases to a patient during ventilation. It provides both an inspiratory and expiratory route. 2. The test lung is an apparatus that contains a breathing bag and and connector. It is connected to the output of a ventilator and is used as part of the device/system check out.. Reason: Reusable latex breathing bags with various part numbers and sold as part of test lung and reusable anaesthesia sets were recalled when Draeger became aware of an incident where a user suffered an acute allergic reaction after using a Dr¿ger latex test lung. The user did not suffer permanent injury. The breathing bag that belongs to the test lung and other latex breathing bags are labelled NR. A.

Recalling firm
Draeger Medical, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-03-25
Terminated
2014-10-08
Firm location
Telford, PA, United States

Product description

Reusable latex breathing bags with multiple part numbers and sold as part of test lung part number and reusable anaesthesia sets. Product Usage: The breathing bag is a component of the anesthesia set (breathing circuit) that is intended to is the conduit for administering medical gases to a patient during ventilation. It provides both an inspiratory and expiratory route. 2. The test lung is an apparatus that contains a breathing bag and and connector. It is connected to the output of a ventilator and is used as part of the device/system check out.

Reason for recall

Reusable latex breathing bags with various part numbers and sold as part of test lung and reusable anaesthesia sets were recalled when Draeger became aware of an incident where a user suffered an acute allergic reaction after using a Dr¿ger latex test lung. The user did not suffer permanent injury. The breathing bag that belongs to the test lung and other latex breathing bags are labelled NR. A

Data sourced from openFDA. This site is unofficial and independent of the FDA.