Recall Z-1448-2013— Philips Healthcare Inc.
Class II · Terminated
Class II FDA medical device recall by Philips Healthcare Inc., initiated 2013-01-16, terminated 2016-09-30. It names one FDA 510(k) clearance: K982795. Product: Philips Diagnost Stationary Radiographic System Models: 712020, 712022, 712050, 712052, 712055, 712057, 712062 Product Usage: Universal RAD Applications. Reason: Installation Acceptance Test Document (IATD) non compliant with regulations. Records shows table height measurements out of specification with the consequence of potential system mis-alignment..
- Recalling firm
- Philips Healthcare Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-01-16
- Terminated
- 2016-09-30
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips Diagnost Stationary Radiographic System Models: 712020, 712022, 712050, 712052, 712055, 712057, 712062 Product Usage: Universal RAD Applications
Reason for recall
Installation Acceptance Test Document (IATD) non compliant with regulations. Records shows table height measurements out of specification with the consequence of potential system mis-alignment.
Data sourced from openFDA. This site is unofficial and independent of the FDA.