Recall Z-1443-2020— RESPIRATORY THERAPEUTICS GROUP LLC

Class II · Ongoing

Class II FDA medical device recall by RESPIRATORY THERAPEUTICS GROUP LLC, initiated 2019-08-19. FDA's recall record names no specific 510(k) clearance for this event. Product: RespVent Closed Suction System T-Piece Tracheostomy Adult, Diameter: 4.6mm (14F), Length: 30.5cm (12in.), MDI Adapter, Cont. 10 Closed Suction Catheter, Sterile R, RxOnly, Reorder Number: RTG-02300, UDI: 00850108006035 - Product Usage: To be used to aspirate liquids or semi solids from a patient s upper airway.. Reason: It has been determined that some units of catheters were not tightly connected to the thumb port. This may result in the catheter disconnecting from the thumb port which could reduce the suction performance and make it difficult to remove from the patient's trachea..

Recalling firm
RESPIRATORY THERAPEUTICS GROUP LLC
Classification
Class II
Status
Ongoing
Initiated
2019-08-19
Firm location
Los Angeles, CA, United States

Product description

RespVent Closed Suction System T-Piece Tracheostomy Adult, Diameter: 4.6mm (14F), Length: 30.5cm (12in.), MDI Adapter, Cont. 10 Closed Suction Catheter, Sterile R, RxOnly, Reorder Number: RTG-02300, UDI: 00850108006035 - Product Usage: To be used to aspirate liquids or semi solids from a patient s upper airway.

Reason for recall

It has been determined that some units of catheters were not tightly connected to the thumb port. This may result in the catheter disconnecting from the thumb port which could reduce the suction performance and make it difficult to remove from the patient's trachea.

Data sourced from openFDA. This site is unofficial and independent of the FDA.