Recall Z-1436-2020— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2020-01-06, terminated 2022-07-11. FDA's recall record names no specific 510(k) clearance for this event. Product: Atellica IM Humidity Pack (Qty 1), SMN 11313496, UDI 00630414243726, Software Version V1.21.0 and lower - Product Usage: Product Usage: The system is intended for the qualitative and quantitative analysis of various body fluids, using photometric, turbidimetric, chemiluminescent, and integrated ion selective electrode technology for clinical use. Humidity packs maintain reagent compartment humidity between 70100%. The operator loads 7 empty humidity packs initially.. Reason: A software error is causing the analyzer to incorrectly eject affected Humidity Packs as expired..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2020-01-06
Terminated
2022-07-11
Firm location
Tarrytown, NY, United States

Product description

Atellica IM Humidity Pack (Qty 1), SMN 11313496, UDI 00630414243726, Software Version V1.21.0 and lower - Product Usage: Product Usage: The system is intended for the qualitative and quantitative analysis of various body fluids, using photometric, turbidimetric, chemiluminescent, and integrated ion selective electrode technology for clinical use. Humidity packs maintain reagent compartment humidity between 70100%. The operator loads 7 empty humidity packs initially.

Reason for recall

A software error is causing the analyzer to incorrectly eject affected Humidity Packs as expired.

Data sourced from openFDA. This site is unofficial and independent of the FDA.