Recall Z-1435-2018— Philips Electronics North America Corporation

Class II · Terminated

Class II FDA medical device recall by Philips Electronics North America Corporation, initiated 2018-03-16, terminated 2020-04-10. It names one FDA 510(k) clearance: K052013. Product: Achieva 1.5T: 1) Nova (Dual), Model 781173, 2) Achieva 1.5T Initial Systems Model 781178, 3) Achieva 1.5T Nova Model 781196, 4) Achieva 1.5T Conversion Model 781283, 5) Achieva 1.5T Model 781296, 6) Achieva 1.5T R5 based Model 781343, 7) Achieva 1.5T Conversion R5 based Model 781346 & 8) Achieva 1.5T Model 781382 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.. Reason: Potential risk for helium gas inside the MR examination room during a magnet quench.

Recalling firm
Philips Electronics North America Corporation
Classification
Class II
Status
Terminated
Initiated
2018-03-16
Terminated
2020-04-10
Firm location
Andover, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Achieva 1.5T: 1) Nova (Dual), Model 781173, 2) Achieva 1.5T Initial Systems Model 781178, 3) Achieva 1.5T Nova Model 781196, 4) Achieva 1.5T Conversion Model 781283, 5) Achieva 1.5T Model 781296, 6) Achieva 1.5T R5 based Model 781343, 7) Achieva 1.5T Conversion R5 based Model 781346 & 8) Achieva 1.5T Model 781382 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.

Reason for recall

Potential risk for helium gas inside the MR examination room during a magnet quench

Data sourced from openFDA. This site is unofficial and independent of the FDA.