Recall Z-1430-2018— Philips Electronics North America Corporation

Class II · Terminated

Class II FDA medical device recall by Philips Electronics North America Corporation, initiated 2018-03-16, terminated 2020-04-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Multiva systems, 1) Multiva 1.5T 8 R5 Model 781072, 2) Multiva 1.5T 16 R5 Model 781073, 3) Multiva 1.5T 8ch R5 based Model 781074, 4) Multiva 1.5T Model 781076, & 5) Multiva 1.5T 16ch R5 based Model 781078 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.. Reason: Potential risk for helium gas inside the MR examination room during a magnet quench.

Recalling firm
Philips Electronics North America Corporation
Classification
Class II
Status
Terminated
Initiated
2018-03-16
Terminated
2020-04-10
Firm location
Andover, MA, United States

Product description

Multiva systems, 1) Multiva 1.5T 8 R5 Model 781072, 2) Multiva 1.5T 16 R5 Model 781073, 3) Multiva 1.5T 8ch R5 based Model 781074, 4) Multiva 1.5T Model 781076, & 5) Multiva 1.5T 16ch R5 based Model 781078 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.

Reason for recall

Potential risk for helium gas inside the MR examination room during a magnet quench