Recall Z-1430-2013— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2013-04-26, terminated 2014-08-04. It names 2 FDA 510(k) clearances: K011720, K003721. Product: AU2700 Clinical Chemistry Analyzer AU5400 Clinical Chemistry Analyzer Perform automated analysis of serum, urine, and other body fluids, including whole blood.. Reason: Beckman Coulter is recalling AU2700/AU2700 Plus & AU5400 Clinical Chemistry Analyzers becuase it was determined that the tubing/joint connections associated with these MODs may leak concentrated detergent. Beckman Coulter has received complaints that the joint connecting the tubing appears to disintegrate over time, causing a leak..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2013-04-26
Terminated
2014-08-04
Firm location
Brea, CA, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

AU2700 Clinical Chemistry Analyzer AU5400 Clinical Chemistry Analyzer Perform automated analysis of serum, urine, and other body fluids, including whole blood.

Reason for recall

Beckman Coulter is recalling AU2700/AU2700 Plus & AU5400 Clinical Chemistry Analyzers becuase it was determined that the tubing/joint connections associated with these MODs may leak concentrated detergent. Beckman Coulter has received complaints that the joint connecting the tubing appears to disintegrate over time, causing a leak.

Data sourced from openFDA. This site is unofficial and independent of the FDA.