Recall Z-1429-2022— Aesculap Implant Systems LLC

Class II · Terminated

Class II FDA medical device recall by Aesculap Implant Systems LLC, initiated 2020-11-18, terminated 2024-12-23. FDA's recall record names no specific 510(k) clearance for this event. Product: ENNOVATE MIS REMOVALKEY SHORT, product code SZ380R. Reason: Fracture or breakage of the spinal fixation arm at the downtube instrument if the 90 degree angle is not met when using the removal key with the spinal fixation system. Redesign of removal key..

Recalling firm
Aesculap Implant Systems LLC
Classification
Class II
Status
Terminated
Initiated
2020-11-18
Terminated
2024-12-23
Firm location
Center Valley, PA, United States

Product description

ENNOVATE MIS REMOVALKEY SHORT, product code SZ380R

Reason for recall

Fracture or breakage of the spinal fixation arm at the downtube instrument if the 90 degree angle is not met when using the removal key with the spinal fixation system. Redesign of removal key.