Recall Z-1429-2020— Seca
Class II · Ongoing
Class II FDA medical device recall by Seca, initiated 2019-06-19. FDA's recall record names no specific 510(k) clearance for this event. Product: Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009. Reason: It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries..
- Recalling firm
- Seca
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2019-06-19
- Firm location
- Hamburg, N/A, Germany
Product description
Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009
Reason for recall
It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries.
Data sourced from openFDA. This site is unofficial and independent of the FDA.