Recall Z-1427-2025— St. Jude Medical

Class II · Ongoing

Class II FDA medical device recall by St. Jude Medical, initiated 2025-02-03. FDA's recall record names no specific 510(k) clearance for this event. Product: CardioMEMS Heart Failure (HF) System that consist of: -CardioMEMS Hospital System Model CM3100 -CardioMEMS Backend Web Application Model CM4000 -Merlin.net Patient Care Network (PCN) Heart Failure Web Application Model MN6000 The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatment.. Reason: Due to issues associated with data migration of patient information and results to a Cloud service, the firm has received complaint associated with some patient profile information being duplicated while other patient information is missing..

Recalling firm
St. Jude Medical
Classification
Class II
Status
Ongoing
Initiated
2025-02-03
Firm location
Atlanta, GA, United States

Product description

CardioMEMS Heart Failure (HF) System that consist of: -CardioMEMS Hospital System Model CM3100 -CardioMEMS Backend Web Application Model CM4000 -Merlin.net Patient Care Network (PCN) Heart Failure Web Application Model MN6000 The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatment.

Reason for recall

Due to issues associated with data migration of patient information and results to a Cloud service, the firm has received complaint associated with some patient profile information being duplicated while other patient information is missing.

Data sourced from openFDA. This site is unofficial and independent of the FDA.