Recall Z-1426-2021— Heartware, Inc.
Class I · Ongoing
Class I FDA medical device recall by Heartware, Inc., initiated 2021-02-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these system components: HVAD Sterile Implant Kit, Model Numbers: 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, MCS1705PU HVAD Controller Kit, Model Numbers: 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1420, 1420JP HVAD AC Adapter Controller, Model Numbers: 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US HVAD DC Adapter Controller: Model Number: 1440 HVAD DATA CABLE 1575- MONITOR, Model Number: 1575 HVAD Battery Pack, Model Numbers: 1650, 1650CA-CLIN, 1650DE HeartWare Patient Pack, Model Number: 1475 HeartWare Waist Pack, Model Numbers: 2050, 2050IL, 2050OUS HeartWare Shoulder Pack, Model Numbers: 2060, 2060IL, 2060OUS The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.. Reason: Medtronic HeartWare HVAD System Instructions for Use (IFU) and Patient Manual (PM) updates regarding: Carrying Cases, Driveline Cover Orientation, and Controller Power Up Sequence issues. The PM is being updated to clarify use of the support strap and wear instructions in addition to cleaning instructions of the carrying cases. The IFU and PM are being updated to add a useful life for the carrying cases. The PM is being updated to inform users to keep the driveline cover on when disconnecting and reconnecting the driveline. The IFU and PM are being updated to clarify the power-up sequence that causes the alarm indicator LEDs and both sets of battery LEDs to turn red for 2.5 seconds while the LCD displays the power-on message. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87603 and 87604.
- Recalling firm
- Heartware, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2021-02-26
- Firm location
- Miami Lakes, FL, United States
Product description
Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these system components: HVAD Sterile Implant Kit, Model Numbers: 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, MCS1705PU HVAD Controller Kit, Model Numbers: 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1420, 1420JP HVAD AC Adapter Controller, Model Numbers: 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US HVAD DC Adapter Controller: Model Number: 1440 HVAD DATA CABLE 1575- MONITOR, Model Number: 1575 HVAD Battery Pack, Model Numbers: 1650, 1650CA-CLIN, 1650DE HeartWare Patient Pack, Model Number: 1475 HeartWare Waist Pack, Model Numbers: 2050, 2050IL, 2050OUS HeartWare Shoulder Pack, Model Numbers: 2060, 2060IL, 2060OUS The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.
Reason for recall
Medtronic HeartWare HVAD System Instructions for Use (IFU) and Patient Manual (PM) updates regarding: Carrying Cases, Driveline Cover Orientation, and Controller Power Up Sequence issues. The PM is being updated to clarify use of the support strap and wear instructions in addition to cleaning instructions of the carrying cases. The IFU and PM are being updated to add a useful life for the carrying cases. The PM is being updated to inform users to keep the driveline cover on when disconnecting and reconnecting the driveline. The IFU and PM are being updated to clarify the power-up sequence that causes the alarm indicator LEDs and both sets of battery LEDs to turn red for 2.5 seconds while the LCD displays the power-on message. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87603 and 87604
Data sourced from openFDA. This site is unofficial and independent of the FDA.