Recall Z-1420-2018— Autochair

Class II · Ongoing

Class II FDA medical device recall by Autochair, initiated 2018-03-07. FDA's recall record names no specific 510(k) clearance for this event. Product: Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B. Reason: Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped..

Recalling firm
Autochair
Classification
Class II
Status
Ongoing
Initiated
2018-03-07
Firm location
Alfreton, United Kingdom

Product description

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Reason for recall

Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.