Recall Z-1419-2026— Medline Industries, LP
Class II · Ongoing
Class II FDA medical device recall by Medline Industries, LP, initiated 2026-01-07. FDA's recall record names no specific 510(k) clearance for this event. Product: Medline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 2) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431K; 3) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431O; 4) KIT ROBOTICS GYN, Model Number: DYKMBNDL109A; 5) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116A; 6) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116G; 7) ROBOTIC PACK, Model Number: DYNJ04135B; 8) MINOR LITHOTOMY PACK, Model Number: DYNJ04382J; 9) NBMC D & C PACK-LF, Model Number: DYNJ0665843C; 10) D&C PACK-LF, Model Number: DYNJ22567G; 11) D&C PACK-LF, Model Number: DYNJ22567J; 12) D AND C, Model Number: DYNJ41272B; 13) D AND C, Model Number: DYNJ41272C; 14) OB PACK, Model Number: DYNJ42884; 15) D&C/GYN PACK, Model Number: DYNJ47713A; 16) PERI GYN PACK, Model Number: DYNJ55377D; 17) ANTERIOR CERVICAL-LF, Model Number: DYNJ58344; 18) D&C PACK, Model Number: DYNJ67214D; 19) D&C PACK, Model Number: DYNJ67214F; 20) D&C PACK, Model Number: DYNJ67214G; 21) D&C PACK, Model Number: DYNJ67214I; 22) LITHOTOMY PACK, Model Number: DYNJ80765C; 23) GYN PACK, Model Number: DYNJ83475; 24) D AND C PACK, Model Number: DYNJ87468; 25) D&C-ASC, Model Number: DYNJ900465C; 26) D&C-ASC, Model Number: DYNJ900465D; 27) D&C-ASC, Model Number: DYNJ900465F; 28) GYN LITHOTOMY-LF, Model Number: DYNJ902711L; 29) GYN LITHOTOMY-LF, Model Number: DYNJ902711N; 30) D&C, Model Number: DYNJ906947C; 31) PACK PERI GYN DSMC, Model Number: DYNJT6424. Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products..
- Recalling firm
- Medline Industries, LP
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-01-07
- Firm location
- Northfield, IL, United States
Product description
Medline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 2) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431K; 3) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431O; 4) KIT ROBOTICS GYN, Model Number: DYKMBNDL109A; 5) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116A; 6) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116G; 7) ROBOTIC PACK, Model Number: DYNJ04135B; 8) MINOR LITHOTOMY PACK, Model Number: DYNJ04382J; 9) NBMC D & C PACK-LF, Model Number: DYNJ0665843C; 10) D&C PACK-LF, Model Number: DYNJ22567G; 11) D&C PACK-LF, Model Number: DYNJ22567J; 12) D AND C, Model Number: DYNJ41272B; 13) D AND C, Model Number: DYNJ41272C; 14) OB PACK, Model Number: DYNJ42884; 15) D&C/GYN PACK, Model Number: DYNJ47713A; 16) PERI GYN PACK, Model Number: DYNJ55377D; 17) ANTERIOR CERVICAL-LF, Model Number: DYNJ58344; 18) D&C PACK, Model Number: DYNJ67214D; 19) D&C PACK, Model Number: DYNJ67214F; 20) D&C PACK, Model Number: DYNJ67214G; 21) D&C PACK, Model Number: DYNJ67214I; 22) LITHOTOMY PACK, Model Number: DYNJ80765C; 23) GYN PACK, Model Number: DYNJ83475; 24) D AND C PACK, Model Number: DYNJ87468; 25) D&C-ASC, Model Number: DYNJ900465C; 26) D&C-ASC, Model Number: DYNJ900465D; 27) D&C-ASC, Model Number: DYNJ900465F; 28) GYN LITHOTOMY-LF, Model Number: DYNJ902711L; 29) GYN LITHOTOMY-LF, Model Number: DYNJ902711N; 30) D&C, Model Number: DYNJ906947C; 31) PACK PERI GYN DSMC, Model Number: DYNJT6424
Reason for recall
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Data sourced from openFDA. This site is unofficial and independent of the FDA.