Recall Z-1416-2021— Becton Dickinson & Co.
Class II · Terminated
Class II FDA medical device recall by Becton Dickinson & Co., initiated 2021-03-24, terminated 2022-08-02. FDA's recall record names no specific 510(k) clearance for this event. Product: BioGX SARS-CoV-2 Reagents for BD MAX System-IVD PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal specimens Catalog Number: 444213. Reason: BioGX SARS-CoV-2 Reagents for BD MAX" System rehydration buffer tubes were not adequately sealed causing some tubes to be partially filled or empty upon use. A partially filled or empty rehydration buffer tube may lead to an Unresolved (UNR) result on the BD MAX" System.
- Recalling firm
- Becton Dickinson & Co.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-03-24
- Terminated
- 2022-08-02
- Firm location
- Sparks, MD, United States
Product description
BioGX SARS-CoV-2 Reagents for BD MAX System-IVD PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal specimens Catalog Number: 444213
Reason for recall
BioGX SARS-CoV-2 Reagents for BD MAX" System rehydration buffer tubes were not adequately sealed causing some tubes to be partially filled or empty upon use. A partially filled or empty rehydration buffer tube may lead to an Unresolved (UNR) result on the BD MAX" System