Class II FDA medical device recall by Becton Dickinson & Co., initiated 2021-03-24, terminated 2022-08-02. FDA's recall record names no specific 510(k) clearance for this event. Product: BioGX SARS-CoV-2 Reagents for BD MAX System-IVD PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal specimens Catalog Number: 444213. Reason: BioGX SARS-CoV-2 Reagents for BD MAX" System rehydration buffer tubes were not adequately sealed causing some tubes to be partially filled or empty upon use. A partially filled or empty rehydration buffer tube may lead to an Unresolved (UNR) result on the BD MAX" System.

Recalling firm
Becton Dickinson & Co.
Classification
Class II
Status
Terminated
Initiated
2021-03-24
Terminated
2022-08-02
Firm location
Sparks, MD, United States

Product description

BioGX SARS-CoV-2 Reagents for BD MAX System-IVD PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal specimens Catalog Number: 444213

Reason for recall

BioGX SARS-CoV-2 Reagents for BD MAX" System rehydration buffer tubes were not adequately sealed causing some tubes to be partially filled or empty upon use. A partially filled or empty rehydration buffer tube may lead to an Unresolved (UNR) result on the BD MAX" System