Recall Z-1413-2017— Encore Medical, Lp
Class II · Terminated
Class II FDA medical device recall by Encore Medical, Lp, initiated 2017-02-09, terminated 2017-10-20. FDA's recall record names no specific 510(k) clearance for this event. Product: RSP Impaction Fixture. Reason: During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating..
- Recalling firm
- Encore Medical, Lp
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-02-09
- Terminated
- 2017-10-20
- Firm location
- Austin, TX, United States
Product description
RSP Impaction Fixture
Reason for recall
During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.