Recall Z-1412-2023— Medtronic Sofamor Danek USA Inc

Class II · Ongoing

Class II FDA medical device recall by Medtronic Sofamor Danek USA Inc, initiated 2023-02-16. It names one FDA 510(k) clearance: K163375. Product: INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.. Reason: Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws.

Recalling firm
Medtronic Sofamor Danek USA Inc
Classification
Class II
Status
Ongoing
Initiated
2023-02-16
Firm location
Memphis, TN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.

Reason for recall

Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

Data sourced from openFDA. This site is unofficial and independent of the FDA.