Recall Z-1412-2017— Encore Medical, Lp

Class II · Terminated

Class II FDA medical device recall by Encore Medical, Lp, initiated 2017-02-09, terminated 2017-10-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Turon Impaction Fixture. Reason: During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating..

Recalling firm
Encore Medical, Lp
Classification
Class II
Status
Terminated
Initiated
2017-02-09
Terminated
2017-10-20
Firm location
Austin, TX, United States

Product description

Turon Impaction Fixture

Reason for recall

During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.