Recall Z-1408-2020— Medtronic, PLC

Class II · Ongoing

Class II FDA medical device recall by Medtronic, PLC, initiated 2020-01-28. FDA's recall record names no specific 510(k) clearance for this event. Product: McGRATH X3 Disposable Laryngoscope Blades-used with the McGRATH MAC Video Laryngoscope to aid the intubation of the trachea Code: X3-002-000 (Individual Unit); X3-003-000 (Carton of 10 Units) - Product Usage: is a tool used to aid the intubation of the trachea. As a rigid laryngoscope it holds and shapes the anatomy allowing a clear view of the larynx and entrance to the trachea.. Reason: Blades do not fit the McGRATH MAC video laryngoscope handle or were tight fitting on the handles which may result in a change in optical performance as displayed on the video screen.

Recalling firm
Medtronic, PLC
Classification
Class II
Status
Ongoing
Initiated
2020-01-28
Firm location
North Haven, CT, United States

Product description

McGRATH X3 Disposable Laryngoscope Blades-used with the McGRATH MAC Video Laryngoscope to aid the intubation of the trachea Code: X3-002-000 (Individual Unit); X3-003-000 (Carton of 10 Units) - Product Usage: is a tool used to aid the intubation of the trachea. As a rigid laryngoscope it holds and shapes the anatomy allowing a clear view of the larynx and entrance to the trachea.

Reason for recall

Blades do not fit the McGRATH MAC video laryngoscope handle or were tight fitting on the handles which may result in a change in optical performance as displayed on the video screen

Data sourced from openFDA. This site is unofficial and independent of the FDA.