Recall Z-1402-2022— Stryker Instruments Div. of Stryker Corporation
Class II · Terminated
Class II FDA medical device recall by Stryker Instruments Div. of Stryker Corporation, initiated 2020-01-22, terminated 2024-01-29. It names one FDA 510(k) clearance: K142801. Product: Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part number 11032). A software application used to help estimation of blood loss. Used with blood bearing canisters and accessories.. Reason: Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in lower than expected results..
- Recalling firm
- Stryker Instruments Div. of Stryker Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-01-22
- Terminated
- 2024-01-29
- Firm location
- Portage, MI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part number 11032). A software application used to help estimation of blood loss. Used with blood bearing canisters and accessories.
Reason for recall
Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in lower than expected results.