Recall Z-1393-2018— GE Medical Systems Ultrasound & Primary Care Diagnostics, LL
Class II · Terminated
Class II FDA medical device recall by GE Medical Systems Ultrasound & Primary Care Diagnostics, LL, initiated 2018-03-16, terminated 2020-12-14. It names one FDA 510(k) clearance: K030962. Product: GE Healthcare Lunar: a) DPX Duo, Model Number: LU41693 b) DPX Bravo, Model Number: LU41692 Provides an estimate of BMD (Bone Marrow Density) at the lumbar spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician.. Reason: Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS..
- Recalling firm
- GE Medical Systems Ultrasound & Primary Care Diagnostics, LL
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-03-16
- Terminated
- 2020-12-14
- Firm location
- Madison, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE Healthcare Lunar: a) DPX Duo, Model Number: LU41693 b) DPX Bravo, Model Number: LU41692 Provides an estimate of BMD (Bone Marrow Density) at the lumbar spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician.
Reason for recall
Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS.
Data sourced from openFDA. This site is unofficial and independent of the FDA.